Prodigy Autocode Bedienungshandbuch Seite 18

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 76
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 17
PRODIGY
AutoCode® Blood Glucose Meter PRODIGY
AutoCode® Blood Glucose Meter
Owner’s Manual Owner’s Manual
34
Caution: To avoid contaminating the control
solution with the content of the test strip,
DO NOT DIRECTLY APPLY CONTROL SOLUTION
ONTO THE TEST STRIP. See Figures 19 and 20.
STEP 5 : Check if the Test Result is in Range.
After seven (7) seconds, the control solution
test result appears on the display. Compare the
test result with the range
printed on the test strip vial.
Each vial of Prodigy
®
No
Coding Test Strips may have
a different control solution
range. The result should fall
within the printed range on
the test strip vial.
See Figure 21.
Figure 19 Figure 20
Figure 21
35
Questionable Results:
If the test result falls outside the range printed
on the test strip vial, check the “Trouble-Shooting
Guide” located in the “System Troubleshooting”
section of this manual and repeat the test.
Results that fall outside of the expected control
ranges may indicate:
• An error in performing the test
• Control solution bottle not shaken well
• Expired or contaminated control solution
• Control solution that is too warm or too cold
• Test strip deterioration
• Meter malfunction
• Test strips expired
Seitenansicht 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 75 76

Kommentare zu diesen Handbüchern

Keine Kommentare