Prodigy Diabetes Care, LLC www.prodigymeter.com
PRODIGY
®
Blood Glucose Meters Owner’s Manual
14
15
Prodigy Diabetes Care, LLC 1.800.243.2636
PRODIGY
®
Blood Glucose Meters Owner’s Manual
Performing a Control Solution Test
Start with the meter off.
STEP 1: Wash your Hands
Wash your hands with mild soap and water before performing any test.
Be sure to dry them thoroughly.
STEP 2: Insert Test Strip
Insert a test strip with the contact bar end entering into the test strip
slotrst.Pushtheteststripasfarasitwillgowithoutbendingit.
The meter turns on automatically and beeps.
STEP 3: Mark as a Control Solution Test
After the “ ” symbol appears on display, press the “M” button and a
“ ” symbol appears on the display. With the “ ” symbol on the display,
the meter will not store your test result in the memory. If you decide not
to perform a control solution test, press the “M” button again and the
“ ” symbol will disappear.
Important: Be sure that you are in control solution mode so
that the test result will not be stored in the meter memory.
STEP 4: Apply Control Solution
• Shakethecontrolsolutionbottle
well, then remove the cap.
• Squeezethebottletodiscardthe
rstdropandwipeoffthedispenser
tip with clean tissue paper or cotton.
• Squeezethebottleagaintogeta
second drop onto a clean,
non-absorbent surface or on
yourngertip.See Figure 3.
• Applythedroptotheopeningofthe
absorbent channel of the test strip (where it meets the narrow
channel)untiltheconrmationwindowislled.See Figures 4
and 5.
• Themeterwillbeepandbegintocountdown.
Caution: To avoid contaminating the control solution with
the content of the test strip, DO NOT DIRECTLY APPLY
CONTROL SOLUTION ONTO THE TEST STRIP.
Figure 3
Figure 4 Figure 5